Emerson 8800 instruction manual Install Tube Fittings

Models: 8800

1 154
Download 154 pages 45.63 Kb
Page 90
Image 90

OCX 8800

Instruction Manual

IM-106-880, Rev 1.0 January 2007

Figure 6-25. Removal

of Tube Fittings

5

3

4

1.

Sensor Housing

 

2.

Electronics Housing

1

3.Tube Fitting

4.Tube Fitting (Type R)

5. Tube Fitting (Type E)

3

2

3

37390070

3

Install Tube Fittings

1.Verify that the replacement tube fitting is identical to the item removed. Special fittings are etched with code markings.

2.Apply pipe thread sealant (Loctite #567) to the mating threads of the tube fitting. Do not apply sealant to the first turn of the external pipe threads.

3.Install and tighten the tube fitting in the mating port of sensor or electronics housing.

6-32