SB Temp Hi (Sample block heater temperature high, > 190ºC)

Probable Cause

Recommended Corrective Action

High noise in OCX power supply

Check power supply for line noise or voltage fluctuations. Install power line filter kit

 

(PN 6A00171G01) or high quality line filter for input power.

O2 Temp Very Hi (Oxygen sensor heater over maximum temperature, > 820ºC)

O2 Htr Rmp Rate (Oxygen sensor heater over maximum temperature ramp rate)

Probable Cause

Recommended Corrective Action

Incorrect O2 heater wiring

Check O2 heater wiring per Figure 7-3and Figure 6-4.Check the wiring at the heater

 

and inside the electronics housing. Correct wiring fault. Perform Reset procedure in

 

Section 3, Configuration and Startup to continue operation.

 

Check O2 thermocouple wiring per Figure 7-3and Figure 6-4.Check the wiring at the

Incorrect O2 thermocouple wiring

 

thermocouple and inside the electronics housing. Correct wiring fault. Perform Reset

 

procedure in Section 3, Configuration and Startup to continue operation.

Electronics package failure

Replace electronics package.

COe Temp Very Hi (Combustion sensor heater over maximum temperature, > 400ºC) COe Htr Rmp Rate (Combustibles sensor heater over maximum temperature ramp rate)

Probable Cause

Recommended Corrective Action

Incorrect COe heater wiring

Check COe heater wiring per Figure 7-3and Figure 6-5.Check the wiring at the heater

 

and inside the electronics housing. Correct wiring fault. Perform Reset procedure in

 

Section 3, Configuration and Startup to continue operation.

Incorrect COe thermocouple wiring

Check COe thermocouple wiring per Figure 7-3and Figure 6-5. Check the wiring at the

 

thermocouple and inside the electronics housing. Correct wiring fault. Perform Reset

 

procedure in Section 3, Configuration and Startup to continue operation.

Electronics package failure

Replace electronics package.

SB Temp Very Hi (Sample block heater over maximum temperature, > 260ºC)

SB Htr Rmp Rate (Sample block heater over maximum temperature ramp rate)

Probable Cause

Recommended Corrective Action

Incorrect sample block heater wiring

Check sample block heater wiring per Figure 7-3and Figure 6-4.Check the wiring at

 

the heater and inside the electronics housing. Correct wiring fault. Perform Reset

 

procedure in Section 3, Configuration and Startup to continue operation.

Incorrect sample block thermocouple wiring

Check sample block thermocouple wiring per Figure 7-3and Figure 6-4. Check the

 

wiring at the thermocouple and inside the electronics housing. Correct wiring fault.

 

Perform Reset procedure in Section 3, Configuration and Startup to continue operation.

Electronics package failure

Replace electronics package.

O2 TC Open (Oxygen sensor heater thermocouple open)

Probable Cause

Recommended Corrective Action

O2 thermocouple or thermocouple circuit open

Check O2 thermocouple and circuit wires for breaks or loose connections per Figure 7-3

 

and Figure 6-4.Repair breaks or loose connections or replace failed thermocouple.

 

Perform Reset procedure in Section 3, Configuration and Startup to continue operation.

COe TC Open (Combustibles sensor heater thermocouple open)

Probable Cause

Recommended Corrective Action

Combustibles sensor thermocouple open

Check thermocouple resistance (lead to ground) of reference and active thermocouples

 

per Figure 7-3and Figure 6-5. If either thermocouple is open or shorted to ground,

 

replace combustibles sensor.

 

 

 

Table continued on next page

7-4

Page 94
Image 94
Emerson 8800 instruction manual SB Temp Hi Sample block heater temperature high, 190ºC

8800 specifications

The Emerson 8800, a pivotal instrument in industrial automation, stands out for its robust features and cutting-edge technologies. Designed to enhance process efficiency and reliability, this device is integral to numerous industries, including oil and gas, chemical, and power generation.

One of the standout features of the Emerson 8800 is its advanced control capabilities. It incorporates a highly flexible control architecture that supports a wide variety of control schemes. This adaptability allows engineers to implement customized solutions tailored to specific process requirements. Additionally, the 8800 series includes integrated predictive diagnostics that continually monitor system performance, alerting operators to potential issues before they escalate into serious problems.

The technology powering the Emerson 8800 is equally impressive. Equipped with state-of-the-art microprocessors, it can handle complex calculations and data processing with remarkable speed and accuracy. The device supports multiple communication protocols, including Foundation Fieldbus, HART, and Modbus. This flexibility ensures seamless integration with existing systems, thereby enhancing data sharing and communication between devices.

Moreover, the Emerson 8800 features a user-friendly interface that simplifies operation and monitoring. The intuitive display allows operators to easily navigate through various settings and real-time data, promoting better decision-making and faster response times. This ergonomic design enhances usability in high-pressure environments, ultimately contributing to improved safety and operational efficiency.

Another key characteristic of the Emerson 8800 is its durability and reliability. Built to withstand the rigors of industrial environments, the device boasts a robust enclosure, ensuring protection against dust, moisture, and extreme temperatures. This reliability minimizes the risk of downtime, making it a cost-effective choice for industries where uptime is critical.

In summary, the Emerson 8800 emerges as a powerhouse in the realm of industrial automation. With its advanced control systems, diverse communication capabilities, user-friendly interface, and rugged design, it embodies efficiency and reliability. As industries continue to evolve, the Emerson 8800 stands ready to meet the challenges of modern automation, driving innovation and productivity in complex environments.