Emerson IM-106-4050 Alarm, Average # 1-4 warning, Average # 1-4 failed, Probe #1-8 failed

Models: IM-106-4050

1 64
Download 64 pages 53.16 Kb
Page 26
Image 26
Alarm

OxyBalance Oxygen Display and Averaging System

Instruction Manual

IM-106-4050 Original Issue March 2006

Alarm

There are two divisions of the Alarm tab: Active Alarms and Alarm History,

 

shown in Figure 3-9and Figure 3-10.

 

There are four possible reports that can be displayed.

 

Average # 1-4 warning

 

This appears when one of the probes fails or is in calibration. When this Alarm

 

appears, the average channel will still display an average O2 % for the

 

remaining probes in the group.

 

Average # 1-4 failed

 

This appears when less than 2 probes in the group are transmitting data for

 

the average calculation. No average percentage will be calculated.

 

Probe # 1-8 calibration in progress

 

This means that the probe is calibrating and will automatically join its specified

 

group when the calibration is complete. Probes in calibration are not factored

 

into the channel average.

 

Probe #1-8 failed

 

This means the probe has failed and needs to be checked. Refer to Section 6,

 

Troubleshooting.

Figure 3-9. Active Alarms

 

3-10

Page 26
Image 26
Emerson IM-106-4050 Alarm, Average # 1-4 warning, Average # 1-4 failed, Probe # 1-8 calibration in progress