OxyBalance Oxygen Display and Averaging System
Instruction Manual
Alarm | There are two divisions of the Alarm tab: Active Alarms and Alarm History, |
| shown in Figure |
| There are four possible reports that can be displayed. |
| Average # |
| This appears when one of the probes fails or is in calibration. When this Alarm |
| appears, the average channel will still display an average O2 % for the |
| remaining probes in the group. |
| Average # |
| This appears when less than 2 probes in the group are transmitting data for |
| the average calculation. No average percentage will be calculated. |
| Probe # |
| This means that the probe is calibrating and will automatically join its specified |
| group when the calibration is complete. Probes in calibration are not factored |
| into the channel average. |
| Probe |
| This means the probe has failed and needs to be checked. Refer to Section 6, |
| Troubleshooting. |