OxyBalance Oxygen Display and Averaging System
Instruction Manual
Averages
The Averages tab of the Current screen, shown in Figure
Fault Condition
A Fault condition is indicated when one or more probes included in an average has failed. The failed probes are not included in the average calculation for that average.
Warning Condition
A warning condition occurs for an average when only all but one channel included in an average is in a failed state. In this state, an "average" cannont be calculated, and the channel will output the value of the remaining functional probe.
Figure