Instruction Manual
Model MicroCEM
a.AC Power
Connect AC power through a 10A circuit breaker that is to be located close to the MicroCEM. The circuit breaker will pro- vide over current protection as well as a means of disconnecting the power.
Maximum power requirements will be 380 watts, with most applications requiring less than this amount.
b.Circular Connector Assembly Instruc- tions
Wire Stripping
Strip insulation from end of wire to be crimped. Do not cut or damage wire strands. Refer to table for proper stripping dimensions.
A
Wire Size | Dim. A |
22O or 22M* | .125 (3.18) |
20 | .188 (4.77) |
16 | .188 (4.77) |
12.188 (4.77)
*Inactive. Not recommended for new design, replacement only.
2.Using correct crimp tool and locator, cycle the tool once to be sure the in- dentors are open. Insert contact and wire into locator. Squeeze tool han- dles firmly and completely to insure a proper crimp. The tool will not re- lease unless the crimp indentors in the tool head have been fully actu- ated.
3.Release crimped contact and wire from tool. Be certain the wire is visi- ble through inspection hole in contact.
Contact Crimping
1.Insert stripped wire into contact crimp pot. Wire must be visible through in- spection hole.
Visual Inspection Hole
Installation | Rosemount Analytical Inc. A Division of Emerson Process Management |