3-3 MicroCEM SETTINGS

The MicroCEM Settings dialog is only available to users with MicroCEM Settings permission. If a user is not currently logged in, the login dialog will be displayed. If the current user doesn’t have MicroCEM Settings permission, a message will be displayed which reads “Permission denied”. When the MicroCEM Settings are invoked from the Tools menu or the MicroCEM Settings button, the MicroCEM Settings tabbed dialog is displayed. The Range page (tab) is displayed initially.

a.Range

The Range Settings page is used to set the range of the Emissions analog out- puts. The outputs support dual range mode. When the emission is below the Range 1 value, the output switches to Range 1 mode and the Range 1 value becomes the full-scale value of the out- put. The range indication digital output will change to the Range 1 state. When the emission is above the Range 1 value, the output switches to Range 2 mode and the Range 2 value becomes the full-scale value of the output. The range indication digital output will change to the Range 2 state. The Range 2 settings cannot be changed and are factory-set.

NOTE

Tabs allow selection of the MicroCEM Settings pages

Figure 3-8. Range Settings

If only one range is needed, set the Range 1 values to the range 2 value. This will disable the dual range fea- ture.

Rosemount Analytical Inc. A Division of Emerson Process Management

Operation

3-7