Model MicroCEM

Instruction Manual

748467-A January 2002

d.Manual Calibration

A dry-run Calibration may be initiated from the Manual Calibration page of the MicroCEM Settings. The results of the calibration will not be applied and only provide a dry run of the calibration. If desired a partial calibration may be in- voked for one or more of the emission types. While the manual calibration is in process, a calibration progress dialog will be displayed as shown in Figure

3-26.When the manual calibration is completed, the results are displayed in the Manual Calibration Results dialog as shown in Figure 3-12.If the Local Calibration checkbox is checked, the Local Calibration valve will be used during the calibration rather than the probe Calibration valve.

NOTE

 

“Start Autocalibration now” will in-

Figure 3-12. Manual Calibration Results

voke a real calibration and will apply

 

new correction factor results when

 

done.

 

Figure 3-11. Manual Calibration

Rosemount Analytical Inc. A Division of Emerson Process Management

Operation

3-9