SECTION 3

OPERATION

CAUTION.

Do not operate or service this instrument before reading and understanding this instruction manual and receiving appropriate training.

Refer to installation drawing supplied with the application data package.

3-1 STARTUP PROCEDURE

Once the MicroCEM has been correctly assembled and installed in accordance with the instructions in Section 2, the analyzer is ready for operation.

Before operating the system, verify that the leak checks have been performed and that the sample handling unit is performing cor- rectly.

Apply power to the system and verify that sample gas is flowing.

NOTE

A warm-up time of from 15 to 50 minutes is required depending on the installed detector(s).

Analyzer operation can be confirmed on the screen of the pocket PC through the glass window on the door. Upon power up, the analyzer will perform a self-test routine. This test will last approximately 60 sec- onds.

3-2 POCKET PC USER INTERFACE

The MicroCEM User Interface runs on a Pocket-PC with Windows CE operating system. It communicates with the Micro- CEM via serial communication port. All input to the Pocket-PC is done using a pointing device that comes with the Pocket- PC.

a.Connecting Pocket PC to MicroCEM

1.Open MicroCEM door. Refer to Figure 3-2.

2.Plug RS232 plug into adapter lo- cated on front panel.

3.Plug power supply cable into 5V adapter

4.Turn Pocket PC on

5.In order to assure no other win- dows are open press the reset button. Reset button is located on the back of the pocket PC.

6.Using the MicroCEM menu click on “programs”

7.Click on “Connection” icon

8.Click on “uCEM” icon.

9.Go to tools menu and click on Mi- croCEM .

10.Unit will display data in 3 to 5 sec- onds.

Figure 3-1. Pocket PC

Rosemount Analytical Inc. A Division of Emerson Process Management

Operation

3-1