Contact Insertion

1.Remove hardware from plug or re- ceptacle and slip over wire bundle in proper order for reassembly.

2.Using proper plastic or metal insertion tool for corresponding contact, posi- tion wire in tip of the tool so that the tool tip butts up against the contact shoulder.

3.Press tool against contact shoulder and, with firm and even pressure, in- sert wired contact and tool tip into center contact cavity. A slight click may be heard as metal retaining tines snap into place behind contact shoul- der.

4.Remove tool and pull back lightly on wire to make sure contact is properly seated. Repeat operation with re- mainder of contacts to be inserted, beginning with the center cavity and working outward in alternating rows.

5.After all contacts are inserted, fill any empty cavities with wire sealing plugs. Reassemble plug or receptacle hard- ware.

Rosemount Analytical Inc. A Division of Emerson Process Management

Installation

2-9