SAFETY INSPECTION/TROUBLESHOOTINGPROCEDURE 2

TROUBLESHOOTING - MECHANICAL

CHAIR VEERS

SLUGGISH TURN/

CASTERS

SQUEAKS AND

LOOSENESS

CHAIR 3

SOLUTIONS

LEFT/RIGHT

PERFORMANCE

FLUTTER

RATTLES

IN CHAIR

WHEELS

 

X

X

X

 

 

X

If pneumatic, check tires for cor-

 

 

 

 

 

 

rect and equal pressure.

X

X

X

X

X

 

Check for loose stem nuts/bolts.

X

 

X

 

 

 

Check that both casters contact

 

 

 

 

 

 

ground at the same time.

 

 

 

 

 

 

 

X

 

 

 

 

 

Check that both stabilizer wheels

 

 

 

 

 

 

are adjusted to the same height.

 

 

 

 

 

 

 

TROUBLESHOOTING - ELECTRICAL

SYMPTOM

PROBABLE CAUSE

SOLUTIONS

Battery won't charge.

Blown Fuse.

Replace fuse (PROCEDURE 6).

 

 

 

Batteries draw excessive current

Battery failure.

Check batteries for shorted cell. Replace if nec-

when charging.

 

essary (BATTERIES-PROCEDURE 11).

 

 

 

Battery indicator flashes the charge

Electrical malfunction.

Contact Dealer/Invacare for Service.

level is low—immediately after re-

Battery failure.

Check batteries for shorted cell. Replace if nec-

charge.

 

essary (BATTERIES-PROCEDURE 11).

 

 

 

Malfunctioning battery charger.

Contact Dealer/Invacare for Service.

 

Electrical malfunction.

Poor connections between charger and wheel-

 

 

chair. Contact Dealer/Invacare for Service.

 

 

 

Battery indicator flashes the charge

Batteries not charged.

Have charger checked.

level is low—too soon after being

Weak batteries.

Replace batteries if necessary (BATTERIES-PRO-

recharged.

 

CEDURE 11).

 

 

 

Electrical malfunction.

Contact Dealer/Invacare for Service.

Motor “chatters” or runs irregular.

Electrical malfunction.

Contact Dealer/Invacare for Service.

Wheelchair does not respond to

One (1) or both motor locks/clutches

Engage motor locks/clutches (PROCEDURE 8).

commands. Power "ON", battery

disengaged.

 

indicator flashes.

 

 

Only one (1) rear wheel turns.

Electrical malfunction.

Contact Dealer/Invacare for Service.

 

One (1) motor lock/clutch is disengaged.

Engage motor lock/clutch (PROCEDURE 8).

 

 

 

Joystick erratic or does not re-

Electrical malfunction.

Contact Dealer/Invacare for Service.

spond as desired.

Controller Programed improperly.

Contact Dealer/Invacare for Service.

 

 

 

 

Wheelchair does not respond to

Poor battery terminal connection.

Clean terminals (BATTERIES-PROCEDURE 11).

commands. Power indicator

 

 

OFF—even after recharging.

Electrical malfunction.

Contact Dealer/Invacare for Service.

 

 

 

 

NOTE: For additional troubleshooting information and explanation of error codes, refer to the individual CONTROL- LER MANUAL supplied with each wheelchair.

CHECKING BATTERY CHARGE LEVEL

The following “Do’s” and “Don’ts” are provided for your convenience and safety.

DON’T

DO

Don’t perform any installation or maintenance

Read and understand this manual and any service information that accompa-

without first reading this manual.

nies a battery and charger before operating the wheelchair.

Don’t perform installation or maintenance of

Move the wheelchair to a work area before checking the fluid level, adding

batteries in an area that could be damaged by

distilled water, cleaning terminals, or opening battery box.

battery spills.

 

Don’t make it a habit to discharge batteries to

Recharge as frequently as possible to maintain a high charge level and extend

the lowest level.

battery life.

 

 

 

 