ELECTRONICS

PROCEDURE 6

REPOSITIONING BATTERY CHARGER CONNECTOR (FIGURE 3)

NOTE: This procedure applies to wheelchairs equipped with the Mark IV controller ONLY.

1.Remove the hex bolt from the T-nut that secures the battery charger connector bracket to the wheelchair.

2.Remove the battery charger connector bracket/ battery charger connector.

3.Slide the T-nut out from the channel on the base frame of the wheelchair.

4.Reposition the T-nut in the channel on the oppo- site side of the wheelchair.

5.Install the hex bolt through the battery charger con- nector bracket AND into the T-nut. Securely tighten.

Hex Bolt

Battery Charger Connector

Battery Charger

Connector Bracket

T-Nut

Channel

Base Frame

NOTE: Base frame shown is for reference ONLY. It is NOT a true representation of the Pronto base frame.

FIGURE 3 - REPOSITIONING BATTERY CHARGER

CONNECTOR

FUSE REPLACEMENT (FIGURE 4)

NOTE: This procedure applies to wheelchairs equipped with the Mark IV controller ONLY.

1.Remove the front shroud. Refer to REMOVING/IN- STALLING THE SHROUDS in PROCEDURE 9 of this manual.

2.Remove the front battery. Refer to REMOVING/IN- STALLING THE BATTERIES in PROCEDURE 11 of this manual.

3.Grasp and pull the existing fuse to remove it from fuse holder.

4.Install new fuse.

5.Reassemble wheelchair by reversing STEPS 1-2.

Fuse

Fuse

Holder

E L E C T R O N I C S

FIGURE 4 - FUSE REPLACEMENT

31