Lifescan Basic Blood Glucose Monitoring System manual 58 mg/dL

Models: Basic Blood Glucose Monitoring System

1 98
Download 98 pages 39.49 Kb
Page 78
Image 78

MESSAGE

PROBLEM

WHAT TO DO

58mg/dL

(Example)

Anotok

redo A

retest

The check strip test result is outside the acceptable range (printed on the back of the meter).

The last check strip test was outside the acceptable range and an acceptable repeat test was not performed.

The last test failed after the countdown started.

Clean the meter and check strip if necessary. Repeat the check strip test.

Repeat the check strip test.

Repeat the test with a new test strip.

70

Page 78
Image 78
Lifescan Basic Blood Glucose Monitoring System manual 58 mg/dL, Last test failed after the countdown started