Note: The control solution range is shown in two different units of
Control solution results
If test results fall outside the range printed on the test strip vial, repeat the test. Results that fall outside the expected range may indicate:
Do not use after expiration date.us | ||||
4 months after first opening. replac | ||||
immediately. |
|
| ||
Date vial first opened __________ | ||||
Lot No. |
| 406814A | ||
Exp. Date |
| 07 - 96 |
| |
ONE TOUCH Normal Control Ran | ||||
de control de normalidad | ||||
86 | - | 122 | MG/DL | |
4 . | 8 | - 6 . 8 | MMOL/L | |
|
| a |
|
|
|
|
|
| |
|
|
|
| |
|
| Milpitas, | California 95035 | |
|
| ©LIFESCAN INC. 1993 AW |
OneTouch® Normal Control Range
(Example)
■ Procedural error |
■ Control solution that has not been |
shaken enough |
■ Expired or contaminated control |
solution |
■ Improper coding of the meter |
Control solution test results are automatically stored with the word contrl. (If you use a logbook, indicate control solution tests by writing “C” after the result.)
■ Dirt on test strip holder or on test |
area beneath the test strip holder |
■ Test strip deterioration |
■ Expired or contaminated test strip |