Important Control Solution Test Information
•Use only OneTouch® Ultra® Control Solution.
•Check the expiration date on the control solution vial. Do not use if expired.
•Control solution, meter, and test strips should come to room temperature
•Shake the vial, discard the first drop of OneTouch® Ultra® Control Solution, and wipe off the dispenser tip to ensure a good sample and an accurate result.
•Use only for three months after first opening. Record the discard date (date opened plus three months) on the control solution vial. Discard after three months.
•Store the control solution tightly closed at room temperature below 30°C (86°F). Do not refrigerate.
CAUTION: The control solution range printed on the test strip vial is for OneTouch® Ultra® Control Solution only. It is used to test meter and test strip performance. It is not a recommended range for your blood glucose level.
13