If the test does not start after applying the sample:

 

PROBABLE CAUSE

WHAT TO DO

 

 

Insufficient blood sample.

Repeat the test with a new

 

 

 

test strip and a larger sample.

 

 

 

 

 

 

Defective test strip.

Repeat the test with a new

 

 

 

test strip.

 

 

Sample applied after automatic

Repeat the test with a new test

 

 

shutoff (two minutes after last

strip; apply sample only when

 

 

user action).

R appears on the display.

 

 

 

 

 

 

Defective meter.

Call 1 800 663-5521.

 

 

 

 

 