If you decide not to perform a control solution test, the “Ë” sign will disappear when you press the C button again.

NOTE: Every time you perform a control solution test, you must mark the test with the “Ë” sign so that the test will be distinguished from a blood glucose test in the meter memory and not included in the 14-day average.

STEP 2 Apply Control Solution.

Shake the control solution vial well. Remove the cap. Squeeze the vial, discard the first drop, and wipe off the dispenser tip to ensure an accurate result. Squeeze the vial again to get a hanging drop. Apply the hanging drop to the TOP WHITE EDGE of the test strip until the drop is drawn onto the test strip, the confirmation window is filled, and the meter begins

to count down.

STEP 3 Result Appears in 15 Seconds.

The meter will count down from “D” to “A” second and the control solution test result will appear. Compare the result with the range printed on the test strip vial. The result should fall within this range.

(Example)

15

Page 25
Image 25
Lifescan OneTouch manual Apply Control Solution