Comparing control solution results
If test results fall outside the range printed on the test strip vial, repeat the test.
the following:
•Error in performing the test.
•Failure to shake the control solution vial vigorously.
•Expired or contaminated control solution.
•Control solution that is too warm or too cold.
•Failure to discard the first drop of control solution and wipe the dispenser tip clean.
•Improper coding of the meter.
•Test strip deterioration.
•Meter malfunction.
CAUTION: If you continue to get control solution test results that fall outside of the range printed on the vial, the system may not be working properly. Do not use the system to test your blood if you continue to get test results that fall outside of the range. If you are unable to resolve the problem, call the OneTouch® Customer Care Line at 1 800
16