STEP 4 Code the Meter.

Move up or down using the Arrow button to the code number that matches the code number on the test strip vial.

The new code number will flash on the display for about three seconds, after which the meter will prompt you to “Apply Blood”.

(Example)

Checking the System

Note: Refer to additional information in the

OneTouch® UltraControl Solution package.

OneTouch® UltraControl Solution is used to check that the meter and the test strips are working together as a system and that you are performing the test correctly.

OneTouch® UltraControl Solution contains a measured amount of glucose that reacts with OneTouch® UltraTest Strips. Compare your control solution test results with the range printed on the test strip vial label. It is very important that you do this simple check routinely to make sure you get accurate results.