MESSAGE
WARNING
Error 4
Strip problem
Please repeat test
See Owner's Booklet
WHAT IT MEANS
1) You may have high glucose and have tested in an environment near the low end of the system’s operating temperature range
2) There may be a problem with the test strip. For example, it may have been damaged or moved during testing.
3) The sample was improperly applied.
ACTION
If you tested in a cool environment, repeat the test in a warmer environment with a new test strip.
If you were testing in a normal or warm environment, repeat the test with a new test strip. (See pages
If you applied the blood incorrectly, review pages
50