MESSAGE

WARNING

Meter is outside of

operating

temperature range

Test not possible

See Owner's Booklet

WARNING

Error 1

Call customer service

See Owner's Booklet

WHAT IT MEANS

The meter has detected that the temperature is outside the system operating range. Do not perform a test until the meter and test strips reach a temperature within the operating range of 6–44°C.

There may be a problem with the meter or the meter may need to be reset.

ACTION

Perform the test after the meter and test strips have reached a temperature within the operating range.

To reset the meter, you must turn the meter off and enter set-up mode (see pages 54- 57). Reset the language, time, and date. You must also reset the meter code to match the code on the test strip vial.

If you continue to get an Error 1 message, do not use the meter. Call your local LifeScan Customer Service office. See back cover for contact information.