Important Control Solution Test Information

Use only OneTouch® UltraControl Solution.

Check the expiration date on the control solution bottle. Record the discard date (date opened plus three months) on the bottle label. Do not use after expiration or discard date, whichever comes first.

Control solution, meter, and test strips should be at room temperature between 20 and 25°C before testing.

Shake the bottle, discard the first drop of control solution, and wipe off the dispenser tip to ensure a good sample and an accurate result.

Store the control solution tightly closed at temperatures below 30°C. Do not refrigerate.

CAUTION: The control solution range printed on the test strip vial is for OneTouch® UltraControl Solution only. It is not a recommended range for your blood glucose level.