
STEP 3 Apply Control Solution.
To ensure an accurate result:
•Shake the bottle well
•Discard the first drop
•Wipe the dispenser tip
Hold the drop to the narrow channel in the TOP EDGE of the test strip. When the confirmation window is full, the meter will begin to count down from 5 to 1.
Note: If you do not apply the control solution within three minutes, the meter will turn itself off. Remove the test strip and insert it back into the test port to restart the test procedure.
STEP 4 Result Appears in 5 Seconds.Compare the control solution result with the range printed on the test strip vial. The result should fall within this range.
(Example)
17