Plumbing
Drainage
^The drainage system is fitted with a
^The machine should preferably be connected to a separate drainage system onsite.
If separate drainage is not available contact your Miele application specialist for advice.
The onsite drain connection point should be located between 1 ft.
(0.3 m) and 3 ft. (1 m) above the lower edge of the machine.
If it is lower than 1 ft. (0.3 m), lay the hose in a curve at a height of at least 1 ft. (0.3 m).
The drainage system must be able to take a minimum drainage flow of
16 l/min.
^The drain hose is approx. 5 ft. (1.5 m)
long, is flexible and has an internal diameter of 7/8 in (22 mm). It must not be shortened. Hose clamps are supplied for securing it in position.
^A longer drain hose (up to 13 ft. [4 m] long) is available to order from the Miele Technical Service Department.
^The drainage system must not exceed 13 ft. (4 meters).
^See the supplied installation diagram.
^Two drain hoses are used, one for the wash cabinet and one for the steam condenser drain.
46