Program Guide

Program Guide

Program

Description

 

 

Custom

Open program available for programming by end user.

 

 

INORGANICA

For removing inorganic residue;

 

generally for analytical applications, for light to moderate soiling, with moderate

 

to high rinse requirements.

 

 

ORGANICA

For removing organic residue, such as oils, fats, wax, agar, preparation

 

applications and for some analytical applications, with moderate to heavy soiling,

 

and moderate rinse requirements. Liquid detergent required, cold, warm, and a

 

DI water supply is recommended.

STANDARD

For a wide range of soiling types that do not denature or precipitate in alkaline

 

environments, for some preparation and analytical applications, for light soiling,

 

and low rinse requirements.

 

 

UNIVERSAL

For removing organic residue (such as proteins, oils and some fats) and some

 

inorganic residues (pH 7, water-soluble metallic salts), for preparation

 

applications and for some analytical applications, moderate to heavy soiling and

 

moderate rinse requirements.

INTENSIVE

For removing organic residue (such as proteins, cell and tissue cultures, oils and

 

some fats) and some inorganic residues (pH 7, water-soluble metallic salts), for

 

preparation applications and analytical applications, moderate to heavy soiling,

 

and moderate to high rinse requirements.

PLASTIC

For temperature-sensitive labware, such as plastic bottles (minimum required

 

temperature resistance 55°C, for preparation applications and for some

 

analytical applications, for light to moderate soiling, for moderate rinse

 

requirements.

 

 

DRAIN

To pump away water, for example when a program is canceled.

 

 

RINSE

To rinse heavily soiled labware, e.g., to pre-treat heavy soiling, remove residue or

 

to avoid drying and encrusting before a full program is run.

 

 

varioTD

For cleaning and thermal disinfection at 93 °C with 5 min holding time (as per

 

EN ISO 15883-1).

 

 

SPECIAL

For cleaning and thermal disinfection at 93 °C with 10 min holding time (as per

93°C - 10'

¶ 18 of Germany's infectious disease control act, section A/B, fungicidal,

bactericidal, and virus-inactivating).

 

 

 

CW = cold water, HW = hot water, DI = distilled water, °F (C°) = Temperature, min = holding time * The DOS 2 rinse aid dosage must be activated by Miele Technical Service