Main menu  (continued)

„„Equipment required for calibration

Konica Minolta CA-310 display color analyzer (optional accessory)

Konica Minolta CA-PU12 or CA-PU15 standard measurement probe (optional accessory)

A RS-232C cable (male to male, straight) (optional accessory)

Connect the RS-232C input terminal on this unit to the RS-232C terminal on the display color analyzer using a straight cable.

Set the Baud rate of the display color analyzer to 38,400.

This unit

Konica Minolta CA-PU15

 

standard measurement probe

RS-232C input

(optional accessory)

 

terminal (9P)

RS-232C cable

Konica Minolta CA-310 display

(optional accessory)

color analyzer (optional accessory)

Analyzer Side

(Straight)

On the monitor

Pin number

Signal

 

 

 

 

 

Pin number

Signal

1

CD

 

 

 

 

 

1

N.C.

2

RXD

 

 

 

 

 

2

TXD

 

 

 

 

 

3

TXD

 

 

 

 

 

3

RXD

 

 

 

 

 

4

DTR

 

 

 

 

 

4

DSR

 

 

 

 

 

5

GND

 

 

 

 

 

5

GND

 

 

 

 

 

6

DSR

 

 

 

 

 

6

DTR

 

 

 

 

 

7

RTS

 

 

 

 

 

7

CTS

 

 

 

 

 

8

CTS

 

 

 

 

 

8

RTS

 

 

 

 

 

9

GND

 

 

 

 

 

9

N.C.

Turn on this unit and perform adequate aging (approx. 1 hour) before starting calibration.

Make the room dark so that no external light can enter the standard measurement probe before starting the calibration. If external light enters the probe, the low brightness characteristics may not be calibrated correctly.

LCD panel characteristics and instrument error in the display color analyzer may sometimes result in small differences in values after calibration.

To further fine tune the monitor, set GAIN and BIAS for R, G and B in the VAR mode of COLOR TEMP.

Do not apply the probe to a WFM/VECTOR display on the screen.

If you are using CA-210, consult the supplier of the unit.

„„CALIBRATION

Select “CALIBRATION” in the “SYSTEM CONFIG” menu to open the following menus.

Sub menu

Settings

Description

AUTO CALIBRATION *8

Connect a CA-310 display color analyzer and use this submenu to make a

 

 

calibration.

 

 

Select “AUTO CALIBRATION” and select “YES” in the confirmation screen that

 

 

appears to start calibration.

RESET *9

Returns calibration data to their factory defaults.

 

 

Select “RESET” and select “YES” in the confirmation screen that appears to return

 

 

calibrated values to their factory defaults.

*8 “EXECUTING” is displayed during “AUTO CALIBRATION” and “COMPLETE” appears when calibration ends. “INCOMPLETE” appears if calibration could not be completed.

*9 When “RESET” ends, “COMPLETE” appears.

Performing AUTO CALIBRATION

1.Turn the rotary knob in the “SYSTEM CONFIG” menu to select the “CALIBRATION” submenu and press the rotary knob.

2.Turn the rotary knob to select the AUTO CALIBRATION submenu and press the rotary knob.

[AUTO CALIBRATION]

YES

NO

MENU EXIT

SEL.

ENTER RETURN RETURN

3.Select “YES” and press the rotary knob. This starts calibration.

[AUTO CALIBRATION]

xxxxxxxxx

MENU EXIT

SEL.

ENTER RETURN RETURN

xxxxxxxxx indicates that one of the following messages with the meaning listed below appears.

EXECUTING: Operation in progress

COMPLETE: Operation completed

INCOMPLETE: Operation incompleted.

(Check the RS-232C connection or display color analyzer connection.)

4.Turn the power off and then on after performing auto cali- bration.

29