4 Operation

Model 3000PA

 

 

 

 

 

 

 

 

 

A A A A A

Retype PWD To Verify

Wait a moment for the entry screen. You will be given clearance to proceed.

A A A A A <ENT> TO Proceed

Use the arrow keys to retype your password and press Enter when finished. Your password will be stored in the microprocessor and the system will immediately switch to the Analyze screen, and you now have access to all instrument functions.

If all alarms are defeated, the Analyze screen appears as:

0.0% Anlz

Range: 0 — 10

If an alarm is tripped, the second line will change to show which alarm

it is:

0.0% Anlz

AL—1

Note: If you log off the system using the logout function in the system menu, you will now be required to re-enter the password to gain access to Span, Zero, Alarm, and Range functions.

4.3.4 Logout

The Logout function provides a convenient means of leaving the analyzer in a password protected mode without having to shut the instrument off. By entering Logout, you effectively log off the instrument leaving the system protected against use until the password is reentered. To log out, press the System button to enter the System function.

Contrast Auto—CalContrast Function is DISABLED

PSWD Logout More

(Refer to Section 1.6)

Use the < > arrow keys to position the blinking over the Logout func- tion, and press Enter to Log out. The screen will display the message:

Protected Until

Password Reentered

4-8

Teledyne Analytical Instruments

Page 40
Image 40