Percent Oxygen Analyzer | Operation 4 | |
|
|
|
|
|
|
|
|
|
4.6The Range Function
The Range function allows the operator to program up to three concen- tration ranges to correlate with the DC analog outputs. If no custom ranges are defined by the user, the instrument defaults to:
Low =
Med =
High =
The Model 3000PA is set at the factory to default to autoranging. In this mode, the microprocessor automatically responds to concentration changes by switching ranges for optimum readout sensitivity. If the current range limits are exceeded, the instrument will automatically shift to the next higher range. If the concentration falls slightly below full scale of the next lower range, the instrument will switch to that range. A corresponding shift in the DC
The autoranging feature can be overridden so that analog output stays on a fixed range regardless of the oxygen concentration detected. If the concentration exceeds the upper limit of the range, the DC output will saturate at 1 V dc.
However, the digital readout and the
4.6.1 Setting the Analog Output Ranges
To set the ranges, enter the range function mode by pressing the
Range button on the front panel.
Use the < > arrow keys to blink the range to be set: low (L), medium (M), or high (H).
Use the ∆∇ arrow keys to enter the upper value of the range (all ranges begin at 0 %). Repeat for each range you want to set. Press Enter to accept the values and return to Analyze mode. (See note below.)
Note: The ranges must be increasing from low to high. For example, if range 1 is set as
Teledyne Analytical Instruments |