A-1 Model 3000PA Specifications

Packaging:

General Purpose

 

• Flush panel mount (Standard).

 

• Rack mount — Relay rack mounted to

 

contain either one or two instruments in one

 

19" relay rack mountable plate (Optional).

Sensor:

Class B-1 Micro-Fuel Cell (standard). Others

 

available.

Cell Block:

Nylon.

90 % Response Time:

10 seconds at 25 °C (77 °F).

Ranges:

Three user definable ranges from 0-1 % to

 

0-100 %, plus air calibration range of 0-25 %.

Alarms:

One system-failure alarm contact to detect

 

power failure.

 

Two adjustable concentration threshold alarms

 

with fully programmable setpoints.

Displays:

2 line by 20 character, alphanumeric, VFD

 

screen.

 

One 5 digit LED display.

Digital Interface:

Full duplex RS-232 communications port.

Power: Universal power supply 85-250 V ac, at

 

47-63 Hz.

Operating Temperature:

0-50 °C (32-122 °F).