Percent Oxygen Analyzer | Operation 4 | |
|
|
|
|
|
|
|
|
|
“C" contacts rated for 3 amperes resistive load at 250 V ac. See Figure in Chapter 3, Installation and/or the Interconnection Diagram included at the back of this manual for relay terminal connections.
The system failure alarm has a fixed configuration described in chapter
3Installation.
The concentration alarms can be configured from the front panel as either high or low alarms by the operator. The alarm modes can be set as latching or
Decide how your alarms should be configured. The choice will depend upon your process. Consider the following four points:
1.Which if any of the alarms are to be high alarms and which if any are to be low alarms?
Setting an alarm as HIGH triggers the alarm when the oxygen concentration rises above the setpoint. Setting an alarm as LOW triggers the alarm when the oxygen concentration falls below the setpoint.
Decide whether you want the alarms to be set as:
•Both high (high and
•One high and one low alarm, or
•Both low (low and
2.Are either or both of the alarms to be configured as failsafe?
In failsafe mode, the alarm relay
3.Are either of the alarms to be latching?
In latching mode, once the alarm or alarms trigger, they will remain in the alarm mode even if process conditions revert back to
4.Are either of the alarms to be defeated?
The defeat alarm mode is incorporated into the alarm circuit so that maintenance can be performed under conditions which would normally activate the alarms.
Teledyne Analytical Instruments |