4 Operation Model 3000PA
4-10 Teledyne Analytical Instruments

4.4 The

Span

Functions

The analyzer is calibrated using span gas.
NOTE: Zero is not necessary for Percent (%) level measurements.
Additional information on Zero functions is provided in the
Appendix A-6 of this manual.
Although the instrument can be spanned using air, a span gas with a
known oxygen concentration in the range of 7090% of full scale of the
range of interest is recommended. Since the oxygen concentration in air is
20.9 %, the cell can take longer to recover if the instrument is used for less
than 1 % oxygen analysis immediately following calibration in air.
Connect the calibration gases to the analyzer according to the instruc-
tions given in Section 3.4.1, Gas Connections, observing all the prescribed
precautions.
Shut off the gas pressure before connecting it to the analyzer, and
be sure to limit the pressure to 40 psig or less when turning it back on.
Readjust the gas pressure into the analyzer until the flowrate (as read on
the analyzers SLPM flowmeter) settles between 0.5 and 2.4 SLPM (ap-
proximately 1-5 scfh).
If you are using password protection, you will need to enter your
password to gain access to either of these functions. Follow the instructions
in section 4.3.3 to enter your password. Once you have gained clearance to
proceed, you can enter the
Zero
or
Span
function.
4.4.1. Cell Failure
When the sensor in the 3000PA begins to fail, the analyzer will usually
require more and more frequent calibration. If the 3000PA analysis readings
drift downward uncharacteristically, try recalibration. If recalibration raises
the readings temporarily, the cell may be failing.
You can check the output of the cell itself by going to the
System
function, selecting More, and pressing
Enter
. The cell output reading will be
on the second line of the display.
Version SelfTest
Cell Output: ### µA
The good reading depends on the class of cell your analyzer is using.