4 Operation

Model 3000PA

 

 

 

 

 

 

 

 

 

The analog output signal has a voltage which depends on the oxygen concentration AND the currently activated analysis range. To relate the signal output to the actual concentration, it is necessary to know what range the instrument is currently on, especially when the analyzer is in the autoranging mode.

To provide an indication of the range, a second pair of analog output terminals are used. They generate a steady preset voltage (or current if you have current outputs) to represent a particular range. The following table gives the range ID output for each analysis range:

Range

Voltage (V)

Current (mA)

LO

0.25

8

MED

0.50

12

HI

0.75

16

CAL (0-25%)

1.00

20

4-18

Teledyne Analytical Instruments

Page 50
Image 50
Teledyne operating instructions Operation Model 3000PA, Range Voltage Current mA