4 Operation Model 3000PA
4-16 Teledyne Analytical Instruments
4.6.2 Autoranging Analysis
Set your analysis ranges as in 4.6.1, above. Leave Mode in Auto, or
use the arrow buttons to change back to Auto.
When operating in autoranging, if the oxygen concentration in your
sample goes ABOVE your HIGHEST range setting, the analyzer will go
into the special 25 % cal range.
However, if one of your range settings is below 0-25 % and another is
set above 0-25 %, the special 0-25 % Air Cal range will NOT activate as the
oxygen level goes through 25 %. Nevertheless, if the oxygen concentration
in your sample goes ABOVE your HIGHEST range setting, the analyzer
will THEN drop back down into the special 25 % cal range.
Once the oxygen concentration drops back down into your highest
range setting, the analyzer will automatically switch back to that range.

CAUTION: While the analyzer is in the Air Cal range, the oxygen

reading cannot go over 25 %, even if the oxygen

concentration is higher than 25 %.

4.6.3 Fixed Range Analysis
The autoranging mode of the instrument can be overridden, forcing the
analyzer DC outputs to stay in a single predetermined range.
To switch from autoranging to fixed range analysis, enter the range
function by pressing the
Range
button on the front panel.
Use the < > arrow keys to move the blinking over AUTO.
Use the ∆∇ arrow keys to switch from AUTO to FX/LO, FX/MED, or
FX/HI to set the instrument on the desired fixed range (low, medium, or
high).
L1.00 M5.00
H10.00 ModeFX/LO
or
L1.00 M5.00
H10.00 ModeFX/MED
or
L1.00 M5.00
H10.00 ModeFX/HI