Percent Oxygen Analyzer

Operation 4

 

 

 

 

 

 

 

 

 

Operation

4.1Introduction

Once the analyzer has been installed, it can be configured for your application. To do this you will:

Set system parameters:

Establish a security password, if desired, requiring Operator to log in.

Establish and start an automatic calibration cycle, if desired.

Calibrate the instrument.

Define the three user selectable analysis ranges. Then choose autoranging or select a fixed range of analysis, as required.

Set alarm setpoints, and modes of alarm operation (latching, failsafe, etc).

Before you configure your 3000PA these default values are in effect:

Ranges: LO = 1 %, MED = 5 %, HI = 10 %

Auto Ranging: ON

Alarm Relays: Defeated, 10 %, HI, Not failsafe, Not latching

Zero: Auto, every 0 days at 0 hours

Span: Auto, at 20.9 %, every 0 days at 0 hours

If you choose not to use password protection, the default password is automatically displayed on the password screen when you start up, and you simply press Enter for access to all functions of the analyzer.

Teledyne Analytical Instruments

4-1

Page 33
Image 33
Teledyne 3000PA operating instructions Percent Oxygen Analyzer Operation, Auto Ranging on