Percent Oxygen Analyzer | Operation 4 | |
|
|
|
|
|
|
|
|
|
Press Escape to
NOTE:When performing analysis on a fixed range, if the oxygen concentration rises above the upper limit (or default value) as established by the operator for that particular range, the output saturates at 1 V dc. However, the digital readout and the
4.7The Analyze Function
Normally, all of the functions automatically switch back to the Analyze function when they have completed their assigned operations. Pressing the Escape button in many cases also switches the analyzer back to the Ana- lyze function. Alternatively, you can press the Analyze button at any time to return to analyzing your sample.
4.8Signal Output
The standard Model 3000PA Percent Oxygen Analyzer is equipped with two
See Rear Panel in Chapter 3, Installation, for illustration.
The signal output for concentration is linear over the currently selected analysis range. For example, if the analyzer is set on range that was defined as
| Voltage Signal | Current Signal |
|
% O2 | Output (V dc) | Output (mA dc) |
|
0 | 0.0 | 4.0 |
|
1 | 0.1 | 5.6 |
|
2 | 0.2 | 7.2 |
|
3 | 0.3 | 8.8 |
|
4 | 0.4 | 10.4 |
|
5 | 0.5 | 12.0 |
|
6 | 0.6 | 13.6 |
|
7 | 0.7 | 15.2 |
|
8 | 0.8 | 16.8 |
|
9 | 0.9 | 18.4 |
|
10 | 1.0 | 20.0 |
|
|
|
| |
| Teledyne Analytical Instruments |