1 Introduction

Model 3000PA

 

 

 

 

 

 

 

 

 

Alarm Connections 2 concentration alarms and 1 system alarm.

RS-232 Port

Serial digital concentration signal output

 

 

and control input.

Remote Probe

Used in the 3000PA for controlling

 

 

external solenoid valves only.

Remote Span/Zero Digital inputs allow external control of analyzer calibration. (See Note, below.)

Calibration Contact To notify external equipment that instrument is being calibrated and readings are not monitoring sample.

Range ID Contacts Four separate, dedicated, range relay contacts. Low, Medium, High, Cal.

Network

For future expansion. Not implemented

 

at this printing.

Optional:

 

Calibration Gas Ports Separate fittings for zero, span and sample gas input, and internal valves for automatically switching the gases.

Current Signal Output Additional isolated 4-20 mA dc plus 4-20 mA dc range ID.

Note: If you require highly accurate Auto-Cal timing, use external Auto-Cal control where possible. The internal clock in the Model 3000PA is accurate to 2-3 %. Accordingly, internally scheduled calibrations can vary 2-3 % per day.

1-6

Teledyne Analytical Instruments

Page 12
Image 12
Teledyne 3000PA operating instructions RS-232 Port, Remote Probe, Network, Optional