Specifications
The analog outputs configured for voltage mode can be calibrated either automatically or manually. In its default mode the instrument is configured for automatic calibration. Note: manual calibration should only be used for the 0.1V range or in cases where the outputs must be closely matched to the characteristics of the recording device. Outputs configured for automatic calibration can be calibrated as a group or individually
Parts list
1.M201E Functional Block Diagram
Moly Test Data Sheet
Maintenance
Configurable Analog Output Overview
Calibration Procedure
Alarms And Cautions
Continue adjustments until the
Page 11
Image 11
Teledyne API Model M201E NH
3
Analyzer Operator Manual
M201E Ammonia Analyzer
Figure
1-2.
M201E and M501 NH
3
, Pneumatic Flow
05206H DCN5910
9
Page 10
Page 12
Page 11
Image 11
Page 10
Page 12
Contents
api-sales@teledyne.com
MANUAL ADDENDUM
MODEL M201E AMMONIA ANALYZER
Page
Name/Description
ABOUT THIS MANUAL
REVISION HISTORY
Part No
05206H DCN5910
This page intentionally left blank
SAFETY MESSAGES
05206H DCN5910
This page intentionally left blank
LIST OF FIGURES
TABLE OF CONTENTS
3.0 CONFIGURABLE ANALOG OUTPUT OVERVIEW
6.0 M201E SPARE PARTS LIST
LIST OF TABLES
1.0 M201E AMMONIA ANALYZER
05206H DCN5910
M201E Ammonia Analyzer
05206H DCN5910
Figure 1-2.M201E and M501 NH3, Pneumatic Flow
M201E Ammonia Analyzer
4NH3 + 5O2 → 4NO + 6H2O
1.1.1Minimizing PMT Drift
→ NO2
1.1.2Purging the Reaction Cell
→ NO
These rules apply also to your calibrator’s internal plumbing. It is highly recommended that you use a M702SS calibrator in conjunction with the M201E. The calibrator has stainless steel plumbing throughout. The M700 calibrator contains Teflon tubing and internal MFC’s that can be affected by the ammonia gas, therefore, the M700 calibrator is not recommend for use with the M201E analyzer. It is important to keep the sampling system well maintained
1.4M201E ANALYZER SPECIFICATIONS
2.0 CALIBRATION PROCEDURE
Comment
Step No
Action
Notes
Step No
Action
Comment
CONV
NO CONV
Exit
Press SPAN, ENTR, EXIT
The M201E enters the calibrate mode from sample
The most important criterion an individual can avail themselves of when spanning an analyzer with ammonia gas is “patience”. Ammonia is a very sticky gas and the response of the analyzer/calibration system depends on a number of factors. The following lists some important points the user should be aware of when calibrating
Press EXIT, EXIT
Table 2-5.Confirming Ammonia Converter Efficiency
TNx CONV
Exit
Press EXIT, EXIT
Generate a known
TNx CONV
Exit
Press CALS
05206H DCN5910
Calibration Procedure
This page intentionally left blank
+ - +
3.0 CONFIGURABLE ANALOG OUTPUT OVERVIEW
Table 3-2.Analog Output Pin-Outs
Table 3-4.Analog Output Current Loop Min/Max
Table 3-3.Analog Output Voltage Range Min/Max
NEXT
ANALOG I / O CONFIGURATION
DIAG
PREV
The analog outputs configured for voltage mode can be calibrated either automatically or manually. In its default mode the instrument is configured for automatic calibration. Note: manual calibration should only be used for the 0.1V range or in cases where the outputs must be closely matched to the characteristics of the recording device. Outputs configured for automatic calibration can be calibrated as a group or individually
ENTR
Configuration
Configurable Analog Output Overview
FROM ANALOG I/O CONFIGURATION MENU
ANALOG I / O CONFIGURATION
The analog outputs in voltage mode can be manually calibrated to closely match the characteristics of the data recorder. Outputs configured for 0.1V full scale should always be calibrated manually
These keys increment/decrement the
Configurable Analog Output Overview
Continue adjustments until the
FROM ANALOG I/O CONFIGURATION MENU
EDIT
3.5ANALOG OUTPUT RECORDER OFFSET ADJUSTMENT
Table 4-1.Preventative Maintenance Schedule
4.0 MAINTENANCE SCHEDULE
CAUTION VERY HOT
Table 4-2.M501 NH3 Converter Rebuild Parts List
Part No
Description
Figure 4-2.Thermocouple Location
2.Remove the cover from the Converter chassis
Table 4-3.Analyzer Flow Check
05206H DCN5910
Maintenance Schedule
This page intentionally left blank
5.0 ALARMS AND CAUTIONS
05206H DCN5910
Figure 5-1.M201E Functional Block Diagram
Alarms and Cautions
Description
6.0 M201E SPARE PARTS LIST
M201E Spare Parts List, Source: PN 06185H DCN5494
Part Number
046030000
OR0000039
05206H DCN5910
M201E Spare Parts List
This page intentionally left blank
NONE
7.0CALCULATING MOLY CONVERTER EFFICIENCIES
05206H DCN5910
Calculating Moly Converter Efficiencies
OUT-GASSING /EATING NO – NOX
MOLY TEST DATA SHEET
LEAK-CHECK WHEN HOT