Teledyne API Model M201E NH3 Analyzer Operator Manual | Calibration Procedure |
Table
Step No. | Action | Comment |
| ||
1. | Press CALS | The M201E enters the calibrate mode from Sample |
|
| mode. The ammonia gas is supplied through the |
|
| SPAN inlet port on the rear panel of the M501 NH3,. |
|
| Calculate the Efficiency factor as outlined in Steps 2 |
|
| in table 2.5. |
2 | Table 2.5 | Follow Steps 4 to 6 in table 2.5. |
|
|
|
Note: The response time of the analyzer increases when using the Zero/Span valve option.