Teledyne API Model M201E NH3 Analyzer Operator Manual

Calibration Procedure

Table 2-7. NH3 Calibration Procedure – NH3 Gas through the SPAN Port

Step No.

Action

Comment

 

1.

Press CALS

The M201E enters the calibrate mode from Sample

 

 

mode. The ammonia gas is supplied through the

 

 

SPAN inlet port on the rear panel of the M501 NH3,.

 

 

Calculate the Efficiency factor as outlined in Steps 2

 

 

in table 2.5.

2

Table 2.5

Follow Steps 4 to 6 in table 2.5.

 

 

 

Note: The response time of the analyzer increases when using the Zero/Span valve option.

05206H DCN5910

23