M201E Ammonia Analyzer | Teledyne API Model M201E NH3 Analyzer Operator Manual |
1.4M201E ANALYZER SPECIFICATIONS
Table
Ranges
Measurement Units
Noise at Zero1
Noise at Span1
Lower Detectable
Limit2
Zero Drift3, 6
Span Drift6
Lag Time
Rise Time4
Fall Time4
Sample Flow Rate
Linearity
Temp Range
Dimensions HxWxD
Weight, Analyzer
Weight, Converter
Weight, Pump
Power, Analyzer
Power, Analyzer5
Power, Pump
Power, Pump CEMark5
Environmental
Analog Output
Analog Output
Resolution
Serial I/O
Status Option
PPB, PPM, microgram/m3, milligram/m3 (user selectable)
<0.5 ppb RMS
<1.0% of reading above 50 ppb
1 ppb RMS
2 ppb / 24 hours
<1.0% FS Range / 24 hours
40 seconds
90% 300 seconds
90% 300 seconds
1000 cm3/min + 10% (500 cm3/min bypass to vacuum manifold, 500 cm3/min to reaction cell)
NO calibration 1% of full scale
NH3 calibration 2% of full scale
7” x 17” x 23.6” (18cm x 43cm x 61cm)
43 lbs (20 kg)
24 lbs (11 kg)
16 lbs (7 kg)
100V ~50/60 Hz, 120V ~60 Hz, 220V ~50Hz, 240V ~50 Hz, 125 watts
230V ~50 Hz, 125 watts
110V ~60 Hz, 220V ~50 Hz, 240V ~50 Hz, 295 watts 230 V ~50 Hz, 2.5 A peak
Installation Category Pollution Degree 2,
0.1, 1V, 5V or 10V. Three outputs convertible to
1 part in 4096 of selected
1x
8 status outputs from
1As defined by USEPA
2Defined as twice the zero noise level
3At constant temperature and voltage
4When pneumatics are conditioned with NH3 overnight and the gases (zero air to NH3) are manually switched at the sample inlet at the rear of the M501.
5Electrical rating for CE Mark Compliance
6Applies when sampling NH3; better results expected for NO/NOx gas measurements.
14 | 05206H DCN5910 |