Teledyne API Model M201E NH3 Analyzer Operator Manual | Configurable Analog Output Overview |
To make these adjustments the auto calibration feature “DATA_OUT_1 AUTO CAL: OFF” must be turned off for each output channel then press the following:
FROM ANALOG I/O CONFIGURATION MENU
DIAG | ANALOG I / O CONFIGURATION |
| |
PREV | NEXT | ENTR | EXIT |
|
|
|
|
Press SET> to select the Analog Output
channel to be configured:
DISPLAYED AS = CHANNEL
DATA_OUT_1 = A1
DATA_OUT_2 = A2
DATA_OUT_3 = A3
DATA_OUT_4 = A4
Then Press EDIT to continue
Press SET Until
DIAG AIO AOUTS CALIBRATED: NO
< SET SET> CAL | EXIT | |
| ||
| ||
DIAG AIO DATA_OUT_1 :5V, TNXCNC1, NO CAL | ||
< SET SET> EDIT | EXIT | |
|
|
|
|
|
|
DIAG AIO | DATA_OUT_1 | RANGE: 5V |
SET> | EDIT | EXIT |
|
| |
|
| |
DIAG AIO | DATA_OUT_1 CALIBRATED: NO | |
< SET | CAL | EXIT |
|
|
|
These keys
increment/decrement the
ZERO/SPAN D-to-A converter
output by
100, 10 or 1 counts
respectively.
Continue adjustments until the
voltage measured at the output of the analyzer and/or the input of the recording device matches the value in the upper right hand corner of the display to the tolerance listed previously.
The analyzer display WILL
NOT CHANGE. Only the
voltage reading of your volt
meter will change.
DIAG AIO DATA_OUT_1 VOLT–Z : 0 mV
U100 UP10 UP DOWN DN10 D100 ENTR EXIT
DIAG AIO | DATA_OUT_1 | |
U100 UP10 UP DOWN DN10 D100 ENTR EXIT | ||
|
| |
|
|
|
DIAG AIO | DATA_OUT_1 CALIBRATED: YES | |
< SET | CAL | EXIT |
|
|
|
EXIT ignores the
new setting.
ENTR accepts the
new setting.