Teledyne M201E Configurable Analog Output Overview, These keys increment/decrement the, Diag

Models: M201E

1 51
Download 51 pages 51.91 Kb
Page 35
Image 35
DATA_OUT_1 = A1

Teledyne API Model M201E NH3 Analyzer Operator Manual

Configurable Analog Output Overview

To make these adjustments the auto calibration feature “DATA_OUT_1 AUTO CAL: OFF” must be turned off for each output channel then press the following:

FROM ANALOG I/O CONFIGURATION MENU

DIAG

ANALOG I / O CONFIGURATION

 

PREV

NEXT

ENTR

EXIT

 

 

 

 

Press SET> to select the Analog Output

channel to be configured:

DISPLAYED AS = CHANNEL

DATA_OUT_1 = A1

DATA_OUT_2 = A2

DATA_OUT_3 = A3

DATA_OUT_4 = A4

Then Press EDIT to continue

Press SET Until

DIAG AIO AOUTS CALIBRATED: NO

< SET SET> CAL

EXIT

 

 

DIAG AIO DATA_OUT_1 :5V, TNXCNC1, NO CAL
< SET SET> EDIT

EXIT

 

 

 

 

 

 

DIAG AIO

DATA_OUT_1
RANGE: 5V

SET>

EDIT

EXIT

 

 

 

 

DIAG AIO

DATA_OUT_1 CALIBRATED: NO
< SET

CAL

EXIT

 

 

 

These keys

increment/decrement the

ZERO/SPAN D-to-A converter

output by

100, 10 or 1 counts

respectively.

Continue adjustments until the

voltage measured at the output of the analyzer and/or the input of the recording device matches the value in the upper right hand corner of the display to the tolerance listed previously.

The analyzer display WILL

NOT CHANGE. Only the

voltage reading of your volt

meter will change.

DIAG AIO DATA_OUT_1 VOLT–Z : 0 mV

U100 UP10 UP DOWN DN10 D100 ENTR EXIT

DIAG AIO

DATA_OUT_1 VOLT–S : 4500 mV

U100 UP10 UP DOWN DN10 D100 ENTR EXIT

 

 

 

 

 

DIAG AIO

DATA_OUT_1 CALIBRATED: YES

< SET

CAL

EXIT

 

 

 

EXIT ignores the

new setting.

ENTR accepts the

new setting.

Figure 3-6. Analog Output Manual Calibration

05206H DCN5910

33

Page 35
Image 35
Teledyne M201E Configurable Analog Output Overview, These keys increment/decrement the, 100, 10 or 1 counts respectively