4.4AUTOZERO FLOW CHECK

Since the M201E is an ammonia analyzer, the flow through the AutoZero orifice is especially important. Check the AutoZero flow as follows:

1.This procedure should be performed with the sample pump running.

2.Remove the top cover of the analyzer. Locate the vacuum manifold at the center rear of the chassis. Locate the 1/8” tube fitting located on the very left side, as viewed from the rear of the analyzer. It will have a label of “0.010” indicating the flow orifice installed.

3.Remove the fitting and attach a calibrated flowmeter to the block fitting capable of measuring in the range of 500 cc/min. The flowmeter should indicate a flow of 500 cc/min ±10%.

4.If the flow is outside this range (most likely lower, due to plugging):

1.Turn off the sample pump and instrument.

2.Remove the 1/8” pipe-to-tube fitting and remove the ¼” sintered filter (p/n FL0000001). This filter is meant to protect the orifice; it will usually become plugged and need replacement rather than the orifice. Replace the filter, and then re-assemble the manifold. Restart the sample pump and recheck the flow.

Table 4-3. Analyzer Flow Check

 

M501 NH3, Converter sample flow check

 

 

Note: Connect both the Teflontubes between analyzer and converter

 

 

GAS PORT

Expected Flow (cc/min)

 

 

 

 

 

 

 

 

“Sample In”

1000 +/- 10 %

 

 

 

 

 

 

 

 

 

 

 

 

 

Analyzer sample flow check

 

 

Note: Disconnect both the Teflontubes between analyzer and converter

 

 

GAS PORT

Expected Flow (cc/min)

 

 

 

 

 

 

“To Conv”

1000 or 500, +/- 10 %

When the analyzer is sampling

0

 

TNx only, the bypass flow is

 

measured. Therefore,

the

flow

 

 

alternates between

500

and

 

 

1000 cc/min. This is the NOx

 

 

channel.

 

 

 

 

 

 

“From Conv”

500 +/- 10%

This is the TNx channel.

 

 

 

 

 

 