4.4AUTOZERO FLOW CHECK
Since the M201E is an ammonia analyzer, the flow through the AutoZero orifice is especially important. Check the AutoZero flow as follows:
1.This procedure should be performed with the sample pump running.
2.Remove the top cover of the analyzer. Locate the vacuum manifold at the center rear of the chassis. Locate the 1/8” tube fitting located on the very left side, as viewed from the rear of the analyzer. It will have a label of “0.010” indicating the flow orifice installed.
3.Remove the fitting and attach a calibrated flowmeter to the block fitting capable of measuring in the range of 500 cc/min. The flowmeter should indicate a flow of 500 cc/min ±10%.
4.If the flow is outside this range (most likely lower, due to plugging):
1.Turn off the sample pump and instrument.
2.Remove the 1/8”
Table 4-3. Analyzer Flow Check
| M501 NH3, Converter sample flow check |
|
| |
Note: Connect both the Teflon™ tubes between analyzer and converter |
|
| ||
GAS PORT | Expected Flow (cc/min) |
|
|
|
|
|
|
|
|
“Sample In” | 1000 +/- 10 % |
|
|
|
|
|
|
|
|
|
|
|
| |
| Analyzer sample flow check |
|
| |
Note: Disconnect both the Teflon™ tubes between analyzer and converter |
|
| ||
GAS PORT | Expected Flow (cc/min) |
|
|
|
|
|
| ||
“To Conv” | 1000 or 500, +/- 10 % | When the analyzer is sampling | ||
0 |
| TNx only, the bypass flow is | ||
| measured. Therefore, | the | flow | |
|
| alternates between | 500 | and |
|
| 1000 cc/min. This is the NOx | ||
|
| channel. |
|
|
|
|
|
| |
“From Conv” | 500 +/- 10% | This is the TNx channel. |
| |
|
|
|
|
|