1.0 M201E AMMONIA ANALYZER

1.0 M201E AMMONIA ANALYZER

This manual is to be used in conjunction with the M200E NOx Analyzer manual, p/n 04410 supplied with this instrument. It is important that you familiarize yourself with the workings of the NOx analyzer before proceeding with NH3 measurements.

The M201E ammonia analyzer consists of a modified M200E nitrogen oxide analyzer and an M501 NH3, ammonia converter. The analyzer measures the following individual gas concentrations: TNx (NH3 and NOx), NOx (NO and NO2) and NO in the sample gas.

The ammonia concentration is calculated by subtracting the NOX reading from the TNx measurement.

The nitrogen dioxide concentration is calculated by subtracting the NO reading from the NOx measurement.

The instrument is available in three pneumatic configurations:

The M201E with External Calibrator Option (ECO) is shown in Figure 1-1.

The basic system is shown in Figure 1-2.

The M201E with Zero/Span valves is shown in Figure 1-3.

Note: For an M201E analyzer with Zero/Span option you must connect the umbilical cord P/N 02255 between the analyzer and the external M501 NH3, converter assembly.

If the analyzer is purchased without the Zero/Span option then the zero and span calibration gases must be individually applied to the sample port. Remember, the gases must be clean/dry and supplied at ambient pressure.

CAUTION

Do not pressurize the Sample/Span/Zero port above ambient