Teledyne M201E manual 5.Confirming Ammonia Converter Efficiency, TNx CONV, Exit, Press EXIT, EXIT

Models: M201E

1 51
Download 51 pages 51.91 Kb
Page 23
Image 23

Teledyne API Model M201E NH3 Analyzer Operator Manual

Calibration Procedure

Table 2-5. Confirming Ammonia Converter Efficiency

Step No.

Action

Comment

 

1.

Press CAL

The M201E enters the calibrate mode from Sample

 

 

mode.

 

2.

Channel Selection

Press TNX then Press ENTR.

 

3.

Range Selection

Press LOW range then Press ENTR.

4.

Press CONC

The following will be displayed:

 

 

 

TNx CONV

Exit

 

 

Press CONV enter 1.0000, Press ENTR

 

 

Press EXIT, EXIT

 

05206H DCN5910