Teledyne API Model M201E NH3 Analyzer Operator Manual

Configurable Analog Output Overview

3.3SELECTING AUTO OR MANUAL ANALOG OUTPUT CALIBRATION

The analog outputs configured for voltage mode can be calibrated either automatically or manually. In its default mode the instrument is configured for automatic calibration. Note: manual calibration should only be used for the 0.1V range or in cases where the outputs must be closely matched to the characteristics of the recording device. Outputs configured for automatic calibration can be calibrated as a group or individually.

To select auto-calibration for a particular channel press the following:

05206H DCN5910