3.5ANALOG OUTPUT RECORDER OFFSET ADJUSTMENT

Configurable Analog Output Overview

Teledyne API Model M201E NH3 Analyzer Operator Manual

3.5ANALOG OUTPUT RECORDER OFFSET ADJUSTMENT

An offset can be added to individual output channels by pressing the following:

FROM ANALOG I/O CONFIGURATION MENU

DIAG

 

ANALOG I / O CONFIGURATION

PREV

NEXT

ENTR

EXIT

 

 

 

 

Press SET Until

These Keys set the Recorder Offset

of the selected

channel

 

DIAG AIO

 

 

AOUTS CALIBRATED: NO

 

 

 

 

SET>

 

 

CAL

 

 

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

DIAG AIO

 

 

DATA_OUT_2: 5V, NH3CNC1, CAL

 

< SET SET>

EDIT

 

 

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

DIAG AIO

 

 

DATA_OUT_2 RANGE: 5V

 

 

 

 

SET>

EDIT

 

 

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

DIAG AIO

 

 

DATA_OUT_2 REC OFS: 0 mV

 

< SET SET>

 

 

EDIT

 

 

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

DIAG AIO

 

 

RECORD OFFSET: 0 MV

 

 

+

0

 

0

0

0

ENTR

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Press SET> to select the Analog Output

channel to be configured:

DISPLAYED AS = CHANNEL

DATA_OUT_1 = A1

DATA_OUT_2 = A2

DATA_OUT_3 = A3

DATA_OUT_4 = A4

Then Press EDIT to continue

Pressing ENTR records the new setting

and returns to the previous menu

Pressing EXIT ignores the new setting and

returns to the previous menu

Figure 3-7. Analog Output Recorder Offset Calibration