Teledyne M201E Diag, Analog I / O Configuration, Prev, Next, Entr, Exit, < Set Set>, Edit, 0.1V

Models: M201E

1 51
Download 51 pages 51.91 Kb
Page 30
Image 30
DIAG

Configurable Analog Output Overview

Teledyne API Model M201E NH3 Analyzer Operator Manual

3.2ANALOG OUTPUT SIGNAL TYPE AND RANGE SELECTION

To select an analog output signal type (DC voltage or current) for one of the analyzer analog output channels press SETUP - MORE -DIAG - 818…ENTER- NEXT…NEXT:

Press SET> to select the channel

parameter:

DATA_OUT_2 RANGE: 5V

DATA_OUT_2 OVERRANGE: ON

DATA_OUT_2 REC OFS: 0 mv

DATA_OUT_2 AUTO CAL: ON

FROM ANALOG I/O CONFIGURATION MENU

DIAG

 

ANALOG I / O CONFIGURATION

 

PREV

NEXT

ENTR

EXIT

 

 

 

 

 

 

DIAG AIO AOUTS CALIBRATED: NO

 

< SET SET>

CAL

EXIT

 

 

 

 

DIAG AIO DATA_OUT_2: 5V, NH3CNC1, CAL

< SET SET>

EDIT

EXIT

 

 

 

 

 

 

 

 

DIAG AIO DATA_OUT_2 RANGE: 5V

 

 

SET>

EDIT

EXIT

 

 

 

 

 

 

 

 

 

 

Press SET> to select the Analog Output channel to be configured:

DISPLAYED AS =

CHANNEL

DATA_OUT_1

=

A1

DATA_OUT_2

=

A2

DATA_OUT_3

=

A3

DATA_OUT_4

=

A4

Then Press EDIT to continue

DATA_OUT_2 CALIBRATED: YES

DATA_OUT_2 OUTPUT: ON

DATA_OUT_2 DATA: NO2CNC1

DATA_OUT_2 SCALE: 500 PPB

DATA_OUT_2 PRECISION: 1

DATA_OUT_2 UPDATE: 5 SEC

Then Press EDIT to continue

DIAG AIO

DATA_OUT_2 RANGE: 5V

 

0.1V 1V

5V 10V CURR
ENTR EXIT

These Keys set the Signal Level of the selected channel

Pressing ENTR records the new setting

and returns to the previous menu

Pressing EXIT ignores the new setting

and returns to the previous menu

Figure 3-2. Analog Output Signal Type