Configurable Analog Output Overview | Teledyne API Model M201E NH3 Analyzer Operator Manual |
3.2ANALOG OUTPUT SIGNAL TYPE AND RANGE SELECTION
To select an analog output signal type (DC voltage or current) for one of the analyzer analog output channels press SETUP - MORE
Press SET> to select the channel
parameter:
DATA_OUT_2 RANGE: 5V
DATA_OUT_2 OVERRANGE: ON
DATA_OUT_2 REC OFS: 0 mv
DATA_OUT_2 AUTO CAL: ON
FROM ANALOG I/O CONFIGURATION MENU
DIAG |
| ANALOG I / O CONFIGURATION |
| |
PREV | NEXT | ENTR | EXIT | |
|
|
| ||
|
|
| ||
DIAG AIO AOUTS CALIBRATED: NO |
| |||
< SET SET> | CAL | EXIT | ||
|
| |||
|
| |||
DIAG AIO DATA_OUT_2: 5V, NH3CNC1, CAL | ||||
< SET SET> | EDIT | EXIT | ||
|
|
| ||
|
|
|
|
|
DIAG AIO DATA_OUT_2 RANGE: 5V |
| |||
| SET> | EDIT | EXIT | |
|
|
|
|
|
|
|
|
|
|
Press SET> to select the Analog Output channel to be configured:
DISPLAYED AS = | CHANNEL | |
DATA_OUT_1 | = | A1 |
DATA_OUT_2 | = | A2 |
DATA_OUT_3 | = | A3 |
DATA_OUT_4 | = | A4 |
Then Press EDIT to continue
DATA_OUT_2 CALIBRATED: YES
DATA_OUT_2 OUTPUT: ON
DATA_OUT_2 DATA: NO2CNC1
DATA_OUT_2 SCALE: 500 PPB
DATA_OUT_2 PRECISION: 1
DATA_OUT_2 UPDATE: 5 SEC
Then Press EDIT to continue
DIAG AIO | DATA_OUT_2 RANGE: 5V |
|
0.1V 1V | 5V 10V CURR | ENTR EXIT |
These Keys set the Signal Level of the selected channel
Pressing ENTR records the new setting
and returns to the previous menu
Pressing EXIT ignores the new setting
and returns to the previous menu
Figure