Teledyne M201E manual Moly Test Data Sheet, Leak-Check When Hot, Out-Gassing /Eating No – Nox

Models: M201E

1 51
Download 51 pages 51.91 Kb
Page 51
Image 51

MOLY TEST DATA SHEET

Line #

TEST

RESULT

 

 

 

2

LEAK-CHECK (WHEN HOT)

YES / NO

3

NOX RESPONSE (MOLY BYPASSED)

__________

4

NOX RESPONSE (MOLY IN-LINE)

__________

5

OUT-GASSING / EATING (NO – NOX)

__________ (>-5 PPB, <5

PPB)

 

 

 

6

(NOx ORIG)

(NOX mode, O3 off)

__________ PPB

7

(NOx REM)

(NOX mode, O3 on)

__________ PPB

8

NOX LOSS (9A - 10B)

__________ (A) (<4% of

NOx ORIG;

 

 

ex: for

450PPB

 

 

 

 

 

4% is 18PPB)

(NO ORIG)

(NO mode, O3 off)

__________ PPB

9

10

(NO REM)

(NO mode, O3 on)

__________ PPB

11

NO2 (9B - 10A)

 

__________ (B)

(>300PPB)

12Efficiency LOSS [ ( A / B ) x 100 ] = [ ( ____A____ / ____B____ ) x 100 ] =

____C____%

13Total Conv Eff [ 100% – C ] = 100% - ____C_____ = _____D_____ % ( > 96%)

05206H DCN5910