: Correct measurement
: Low in reliability
: Incorrect measurement
: Measurement value exceeding the measuring range

Display of Measurement Values

Measurement values are displayed on the monitor screen for up to three measure- ments. From the fourth measurement on, values of earlier measurements are deleted in order.

Figure only Figure in ( )

ERR

OVER

If the result is a figure in parentheses or ERR, do the measurement again, making sure the patient does not blink and eyelashes do not get in the outer alignment mark. If OVER is displayed, switch the measuring range to 0-60 and do the measurement again.

MEASUREMENT UNDER MANUAL MODE

MEMO

Adjust the height of the automatic instrument table so that correct measurement values can be obtained by allowing the patient to un- dergo measurements in a comfortable position.

Make the patient relaxed so as to secure correct measurement val- ues; make sure the patient does not hold his breath or remain tense.

Setting the Measurement Mode

The initial status of the measurement mode is AUTO, upon turning the power ON.

1

Return to the Measurement screen.

2

Press

on the control panel and change the measurement mode display to

 

“MANU”.

 

Setting the Measuring Range

See page 33.

Alignment and Measurement

The alignment operation is controlled through the control lever.

For details about the adjustment of the machine body using the control lever, see “Memo” on page 34.

38

BASIC OPERATIONS

Page 39
Image 39