Topcon CT-80, COMPUTERIZED TONOMETER Before Requesting Service, Checking Operations

Models: COMPUTERIZED TONOMETER CT-80

1 59
Download 59 pages 44.69 Kb
Page 46
Image 46

BEFORE REQUESTING SERVICE

CHECKING OPERATIONS

Air Check

If a problem is suspected, do an air check.

If the result is “NG (+)” or “NG (-),” call your dealer.

For instructions on how to perform an air check, see “Air Check” on page 29.

Checking Operations

WARNING

To avoid electric shocks, do not open the instrument.

Refer all servicing to qualified personnel.

To avoid electric shocks, do not remove the covers from WARNING the bottom and top surfaces, TV monitor, measuring unit,

etc.

If a problem is suspected, perform checks following the Check List shown below. If the condition is not improved by the suggested remedy or if it is not described in the list, call your dealer.

CHECK LIST

Problem:

Check point:

Remedy:

Page:

Monitor screen does not work.

Power cable is not plugged into

Secure plug in power cable.

P.14

 

receptacle.

 

 

 

 

 

Power cable is not plugged into

Plug power cable into machine

P.14

 

machine body.

body.

 

 

 

Power save function is on.

Return to normal status.

P.28

 

Fuse is burned.

Replace fuse.

P.55

Monitor screen display is not

Monitor screen needs to be

Adjust monitor screen.

P.54

clear.

readjusted.

 

 

 

 

Measuring window needs to be

Clean measuring window.

P.52

 

cleaned.

 

 

 

 

Auto mode measurement is not

Measuring window needs to be

Clean measuring window.

P.52

possible.

cleaned.

 

 

 

 

Window in nozzle needs to be

Clean window in nozzle.

P.53

 

cleaned.

 

 

 

 

 

Condition of patient's eye is

Measure under Manual mode.

P.38

 

unfavorable.

 

 

 

 

 

Manual mode is on.

Set Auto mode.

P.33

Measurement values have ( ) or

Measuring window is not clear

Clean window in nozzle.

P.53

ERRs are displayed.

enough.

 

 

 

Paper comes out unprinted.

Printer paper winding is re-

Set printer paper correctly.

P.23,26

 

versed.

 

 

 

 

Paper does not come out.

Printer paper is used up.

Supply printer paper.

P.23, 56

 

Paper is jammed.

Remove jammed paper.

P.55

Machine body does not move.

Clamping knob/fixing knob is

Loosen up clamping knob/fixing

P.8

 

fastened.

knob.

 

 

45

BEFORE REQUESTING SERVICE

Page 46
Image 46
Topcon CT-80, COMPUTERIZED TONOMETER instruction manual Before Requesting Service, Checking Operations