Replace the fuse with the attached fuse.

Replacing the Fuse

Press the fuse holder with a screwdriver and turn it clockwise. The fuse holder is now reset.

Setting the Fuse Holder

SPECIAL NOTES ON CLEANING

Cleaning the Outer Cover

CAUTION

Do not use or apply any spray-typed cleaner near the in- strument. If a drop of cleaner remains inside the measur- ing nozzle, the patient’s eye may be injured during meas- urement.

MEMO

To avoid discoloring/deterioration of the plastic components, do

not use volatile solvents (benzine, thinner, gasoline, etc.).

When the outer cover, operation panel and the like become stained, clean them with a dry cloth. If the stain is extreme, a neutral detergent for tableware may be used by diluting it with tepid water. Soak the cloth in the solution, squeeze it out, and then use if to wipe off stains.

56

MAINTENANCE AND CHECKING

Page 57
Image 57
Topcon COMPUTERIZED TONOMETER, CT-80 instruction manual Special Notes on Cleaning, Cleaning the Outer Cover