6. Maintenance

6. Maintenance

6. Maintenance

6.1 General

Warning! Always disconnect MultiTemp III from the mains

 

(Pull out the mains plug) before maintenance and/or service.

 

Service should be carried out by authorised personnel only.

 

To maintain efficient operation of the compressor cooling

 

unit, have the air grid and liquifier (heat exchanger) checked

 

and cleaned by service personnel two to four times a year,

 

depending on the amount of dust in the air.

 

Always ensure the liquid bath has sufficient circulating liqu-

 

id.

 

The cooling coil in the bath and the bath vessel itself should

 

be periodically checked and any deposits removed. Metal

 

particles and acidic or alkaline substances should be removed

 

quickly as they could harm the stainless steel material and

 

cause corrosion.

 

After some time, the stainless steel surface and the housing

 

may show spots and become tarnished. Clean them with

 

stainless steel liquid cleaners normally used in the kitchen.

 

Do not use scouring pads or powder!

 

Disposal

 

MultiTemp III will provide many years of efficient operation.

 

When its working life is over, however, please remember that

 

the unit contains coolant R 134a, which despite its environ-

 

mental friendliness may only be discarded by authorised per-

 

sonnel.